GOST 33729-2016 GOST 20996.3-2016 GOST 31921-2012 GOST 33730-2016 GOST 12342-2015 GOST 19738-2015 GOST 28595-2015 GOST 28058-2015 GOST 20996.11-2015 GOST 9816.5-2014 GOST 20996.12-2014 GOST 20996.7-2014 GOST P 56306-2014 GOST P 56308-2014 GOST 20996.1-2014 GOST 20996.2-2014 GOST 20996.0-2014 GOST 16273.1-2014 GOST 9816.0-2014 GOST 9816.4-2014 GOST P 56142-2014 STATE STANDARD P 54493-2011 GOST 13498-2010 STATE STANDARD P 54335-2011 GOST 13462-2010 STATE STANDARD P 54313-2011 GOST P 53372-2009 GOST P 53197-2008 GOST P 53196-2008 GOST P 52955-2008 GOST P 50429.9-92 GOST 6836-2002 GOST 6835-2002 GOST 18337-95 GOST 13637.9-93 GOST 13637.8-93 GOST 13637.7-93 GOST 13637.6-93 GOST 13637.5-93 GOST 13637.4-93 GOST 13637.3-93 GOST 13637.2-93 GOST 13637.1-93 GOST 13637.0-93 STATE STANDARD 13099-2006 GOST 13098-2006 GOST 10297-94 GOST 12562.1-82 GOST 12564.2-83 GOST 16321.2-70 GOST 4658-73 GOST 12227.1-76 GOST 16274.0-77 GOST 16274.1-77 GOST 22519.5-77 GOST 22720.4-77 GOST 22519.4-77 GOST 22720.2-77 GOST 22519.6-77 GOST 13462-79 GOST 23862.24-79 GOST 23862.35-79 GOST 23862.15-79 GOST 23862.29-79 GOST 24392-80 GOST 20997.5-81 GOST 24977.1-81 GOST 25278.8-82 GOST 20996.11-82 GOST 25278.5-82 GOST 1367.7-83 GOST 26239.9-84 GOST 26473.1-85 GOST 16273.1-85 GOST 26473.2-85 GOST 26473.6-85 GOST 25278.15-87 GOST 12223.1-76 GOST 12645.7-77 GOST 12645.1-77 GOST 12645.6-77 GOST 22720.3-77 GOST 12645.4-77 GOST 22519.7-77 GOST 22519.2-77 GOST 22519.0-77 GOST 12645.5-77 GOST 22517-77 GOST 12645.2-77 GOST 16274.9-77 GOST 16274.5-77 GOST 22720.0-77 GOST 22519.3-77 GOST 12560.1-78 GOST 12558.1-78 GOST 12561.2-78 GOST 12228.2-78 GOST 18385.4-79 GOST 23862.30-79 GOST 18385.3-79 GOST 23862.6-79 GOST 23862.0-79 GOST 23685-79 GOST 23862.31-79 GOST 23862.18-79 GOST 23862.7-79 GOST 23862.1-79 GOST 23862.20-79 GOST 23862.26-79 GOST 23862.23-79 GOST 23862.33-79 GOST 23862.10-79 GOST 23862.8-79 GOST 23862.2-79 GOST 23862.9-79 GOST 23862.12-79 GOST 23862.13-79 GOST 23862.14-79 GOST 12225-80 GOST 16099-80 GOST 16153-80 GOST 20997.2-81 GOST 20997.3-81 GOST 24977.2-81 GOST 24977.3-81 GOST 20996.4-82 GOST 14338.2-82 GOST 25278.10-82 GOST 20996.7-82 GOST 25278.4-82 GOST 12556.1-82 GOST 14339.1-82 GOST 25278.9-82 GOST 25278.1-82 GOST 20996.9-82 GOST 12554.1-83 GOST 1367.4-83 GOST 12555.1-83 GOST 1367.6-83 GOST 1367.3-83 GOST 1367.9-83 GOST 1367.10-83 GOST 12554.2-83 GOST 26239.4-84 GOST 9816.2-84 GOST 26473.9-85 GOST 26473.0-85 GOST 12645.11-86 GOST 12645.12-86 GOST 8775.3-87 GOST 27973.0-88 GOST 18904.8-89 GOST 18904.6-89 GOST 18385.0-89 GOST 14339.5-91 GOST 14339.3-91 GOST 29103-91 GOST 16321.1-70 GOST 16883.2-71 GOST 16882.1-71 GOST 12223.0-76 GOST 12552.2-77 GOST 12645.3-77 GOST 16274.2-77 GOST 16274.10-77 GOST 12552.1-77 GOST 22720.1-77 GOST 16274.4-77 GOST 16274.7-77 GOST 12228.1-78 GOST 12561.1-78 GOST 12558.2-78 GOST 12224.1-78 GOST 23862.22-79 GOST 23862.21-79 GOST 23687.2-79 GOST 23862.25-79 GOST 23862.19-79 GOST 23862.4-79 GOST 18385.1-79 GOST 23687.1-79 GOST 23862.34-79 GOST 23862.17-79 GOST 23862.27-79 GOST 17614-80 GOST 12340-81 GOST 31291-2005 GOST 20997.1-81 GOST 20997.4-81 GOST 20996.2-82 GOST 12551.2-82 GOST 12559.1-82 GOST 1089-82 GOST 12550.1-82 GOST 20996.5-82 GOST 20996.3-82 GOST 12550.2-82 GOST 20996.8-82 GOST 14338.4-82 GOST 25278.12-82 GOST 25278.11-82 GOST 12551.1-82 GOST 25278.3-82 GOST 20996.6-82 GOST 25278.6-82 GOST 14338.1-82 GOST 14339.4-82 GOST 20996.10-82 GOST 20996.1-82 GOST 12645.9-83 GOST 12563.2-83 GOST 19709.1-83 GOST 1367.11-83 GOST 1367.0-83 GOST 19709.2-83 GOST 12645.0-83 GOST 12555.2-83 GOST 1367.1-83 GOST 9816.3-84 GOST 9816.4-84 GOST 9816.1-84 GOST 9816.0-84 GOST 26468-85 GOST 26473.11-85 GOST 26473.12-85 GOST 26473.5-85 GOST 26473.7-85 GOST 16273.0-85 GOST 26473.3-85 GOST 26473.8-85 GOST 26473.13-85 GOST 25278.13-87 GOST 25278.14-87 GOST 8775.1-87 GOST 25278.17-87 GOST 18904.1-89 GOST 18904.0-89 STATE STANDARD P 51572-2000 GOST 14316-91 GOST P 51704-2001 GOST 16883.1-71 GOST 16882.2-71 GOST 16883.3-71 GOST 8774-75 GOST 12227.0-76 GOST 12797-77 GOST 16274.3-77 GOST 12553.1-77 GOST 12553.2-77 GOST 16274.6-77 GOST 22519.1-77 GOST 16274.8-77 GOST 12560.2-78 GOST 23862.11-79 GOST 23862.36-79 GOST 23862.3-79 GOST 23862.5-79 GOST 18385.2-79 GOST 23862.28-79 GOST 16100-79 GOST 23862.16-79 GOST 23862.32-79 GOST 20997.0-81 GOST 14339.2-82 GOST 12562.2-82 GOST 25278.7-82 GOST 20996.12-82 GOST 12645.8-82 GOST 20996.0-82 GOST 12556.2-82 GOST 25278.2-82 GOST 12564.1-83 GOST 1367.5-83 GOST 25948-83 GOST 1367.8-83 GOST 1367.2-83 GOST 12563.1-83 GOST 9816.5-84 GOST 26473.4-85 GOST 26473.10-85 GOST 12645.10-86 GOST 8775.2-87 GOST 25278.16-87 GOST 8775.0-87 GOST 8775.4-87 GOST 12645.13-87 GOST 27973.3-88 GOST 27973.1-88 GOST 27973.2-88 GOST 18385.6-89 GOST 18385.7-89 GOST 28058-89 GOST 18385.5-89 GOST 10928-90 GOST 14338.3-91 GOST 10298-79 GOST P 51784-2001 GOST 15527-2004 GOST 28595-90 GOST 28353.1-89 GOST 28353.0-89 GOST 28353.2-89 GOST 28353.3-89 STATE STANDARD P 52599-2006

GOST 12555.2−83 Alloys, silver-platinum. Method of spectral analysis (with Change No. 1)


GOST 12555.2−83

Group B59


INTERSTATE STANDARD


ALLOYS OF SILVER AND PLATINUM

Method of spectral analysis

Silver-platinum alloys. Methods of spectral analysis


AXTU 1709

Date of introduction 1984−07−01


INFORMATION DATA

1. DEVELOPED AND INTRODUCED by the Ministry of nonferrous metallurgy of the USSR

DEVELOPERS

A. A. Kuranov, G. S. HAK, N. With. Stepanova, N. D. Sergienko, T. I. Belyaeva, I. G., Sazhin

2. APPROVED AND put INTO EFFECT by Decision of the USSR State Committee for standards from 25.03.83 N 1371

3. REPLACE GOST 12555−67 in part of sec. 3

4. REFERENCE NORMATIVE AND TECHNICAL DOCUMENTS

   
The designation of the reference document referenced
Item number
GOST 3118−77
Sec. 2
GOST 22864−83
1.1

5. Limitation of actions taken by Protocol No. 3−93 Interstate Council for standardization, Metrology and certification (ICS 5−6-93)

6. The re-release (Feb 1999) with amendment No. 1, approved in September 1988 (IUS 1−89)


This standard specifies a method of determining the spectral impurities iridium, palladium, rhodium, gold, iron (at mass proportion of each from 0.01 to 0.10%) and lead (for the mass concentration of from 0.001 to 0.020%) in the silver-platinum alloys.

The method is based on the translation of alloy into the globule (liquid drop of melt) and measuring the intensity of the impurity lines in the arc spectrum. Quantification of the mass fraction of impurities is set by calibration with standard samples.

1. GENERAL REQUIREMENTS

1.1. General requirements for method of analysis according to GOST 22864.

1.2. The numerical value of the result of the analysis must end with the digit the same category, and permissible differences.

(Added, Rev. N 1).

2. APPARATUS, REAGENTS AND SOLUTIONS


Spectrograph medium dispersion.

Arc generator of alternating current.

The three-stage attenuator.

Microphotometer.

Standard samples of the enterprise.

Electrodes, carbon, made from coal spectrality, brand high purity-7.

The lathe table for sharpening carbon electrodes.

Photographic plates of the spectral type II sensitivity of 10−15 conventional units.

The developer and fixer.

Hydrochloric acid according to GOST 3118, diluted 1:1.

(Changed edition, Rev. N 1).

3. PREPARATION FOR ASSAY


Samples for analysis must be in the form of strips or chips. To remove surface contamination samples before taking the test portions are boiled in hydrochloric acid (1:1) for 2 min, then washed with water and dried.

The electrodes were applied to the spectral coals with a diameter of 6 mm. the Lower electrode is notched with millimeter recess in the end face, in which is placed the sample to be analyzed or of standard sample mass of 0.10 g. the Counter serves as a carbon rod, sharpened to a hemisphere or a truncated cone with ground diameter of 1.5−2 mm.

(Changed edition, Rev. N 1).

4. ANALYSIS


Spectra photographed at slit width of the spectrograph 0.025 mm, exposure 30, a current of 5 A through a three-stage attenuator. The distance between electrodes 2.5 mm is adjusted during the exposure on the intermediate diaphragm. The photographic plate is placed in the cassette, departing 4 cm from the right far edge of the cassette.

For each of the analyzed and the standard sample receive three parallel definitions.

Photographic plates are in the developer temperature of 18−20 °C for 6 min.

Shown photographic plate is then rinsed with water, fixed, washed in running water for 10 min, dried and photometric.

5. PROCESSING OF THE RESULTS

5.1. Wavelength of analytical spectral lines is given in the table.

       
The designated element
Wavelength of lines of the designated element, nm
Element comparison
The wavelength of the line item comparisons, nm
Palladium
342,12
Platinum
315,66
Iridium
313,33
The same
315,66
Rhodium
343,49
«
315,66
Gold
267,59
«
280,32
Iron
259,94
«
280,32
Lead
283,31
«
 

5.2. Determination of the mass fraction of the elements are by the method of «three standards» with the objective photoretrieval. The calibration graphs are built for each analyzed element. Y-axis delay value difference pucherani lines of the designated element, and the abscissa shows the values of the logarithm of mass fractions of standard samples.

With the help of calibration curve by the known values of the difference of pochernenija find the contents of the analyzed element in the sample.

5.3. The convergence of the method is characterized by the relative standard deviation of 0.15.

The final result of the analysis be the arithmetic mean of three parallel measurements under the condition of:

ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1),


where ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1)is the greatest result of the parallel measurements;

ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1)least the result of the parallel measurements;

ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1) — relative standard deviation describing repeatability of measurements;

ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1) — the arithmetic mean calculated from ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1)parallel measurements (ГОСТ 12555.2-83 Сплавы серебряно-платиновые. Метод спектрального анализа (с Изменением N 1)3).